SKU : QL-800-O³
Gas detected: ozone (O³) Detection principle: Electrochemical principle Sampling method: pump suction type Measure Range: 0-100ppm Resolution: 0.01ppm Precision: 2%FS Response time: ≤10S Repeatability: ≤±1% Zero shift: ≤±1% (F.S/year)
หมวดหมู่ : 3.ระบบเกลือ และระบบฆ่าเชื้ออื่นๆ ,  3.2 ระบบโอโซน ,  winwinpool ,  Ozone Testing Meter ,  เครื่องผลิตโอโซน , 
แบรนด์ : winwinpool
Share
เครื่องวัดโอโซน (Portable Ozone Gas Sensor)
รุ่น QL-800-O³
QL-800-O3 is an essential equipment of continuous detection of ozone gas concentration,suitable for indoor ozone sterilization, ozone generator outlet concentration, can accurately detect the ozone density. With secondary light alarm function, greatly satisfy the scene of the industrial safety monitoring of equipment reliability requirement, this product is widely used in petroleum chemical industry, industrial production,flue gas,exhaust gas environment, biological pharmaceuticals, environmental protection, school laboratory,etc.
Product Features
♦ Imported high precision electrochemical sensors
♦ USB high speed data transmission,data downloadable(optional))
♦ LCD dot matrix display technology ,displace gas type,gas unit,gas density,maximum test value,local time,ambient humidity
♦ Chinese/English control interface
♦ Inner sample pump
♦ Explosion-proof conformity certification, Exia II CT4
Product Function
♦Monitoring of the environment or confined space gas concentrations and alarms with self check function.
♦Zero calibration feature, the gas monitoring is more accurate, more reliable.
♦A key to restore the factory settings function, can be removed from misuse problems.
♦With temperature and pressure compensation,perfect realization of compensation under different temperature and pressure environment .
♦Two alarm modes: sound and light alarm.
♦High-capacity lithium polymer rechargeable battery.
♦Pump suction sampling type.
♦The shell is made of special high-strength plastic, non-slip, waterproof, dustproof, explosion
Exquisite high-end aluminum suitcase.
Technical parameters
Gas Detected | Ozone(O3) |
Detection principle | Electrochemical principle |
Sampling Method | Pump suction type, |
Measure Range | 0-100ppm/0-10ppm/0- 200ppm/0-500ppm/0- 1000ppm |
Resolution | 0.01ppm |
Precision | 2%FS |
Response Time | ≤10S |
Repeatability | ≤±1% |
Zero shift | ≤±1%(F.S/year) |
Operating language | Chinese/English |
Temperature | -40~70 |
Recovery Time | ≤10S |
Linearity error | ≤±1% |
Display | Digital LCD backlit display |
Gas unit | multiple gas units to switch the display |
Humidity | 0-95%RH |
Degree of protection | IP66 |
Operating time | 100 Hr |
Weight | 300g |
Explosion-Proof | ExibCT4 |
Alarm mode | Sound, light, vibration |
Dimensions | 205*75*32mm |